The HA1500 series analyzer is particularly suitable for measuring the Vol% level of a gas in binary or pseudo binary mixtures.
The HA1500 series analyzer is particularly suitable for measuring the% level of a gas in binary or pseudo binary mixtures. For example, hydrogen (H2), helium (He), dissociated ammonia, sulfur hexafluoride (SF6), and inert gases such as krypton (Kr) and xenon (Xe) can all be measured using this technology. The thermal conductivity method is particularly suitable for measuring gases without dedicated sensors and can provide a cost-effective analysis solution.
According to the process conditions, different sample gas pretreatment systems (standard or customized) can also be selected. For applications with corrosive gas environments, different sensor devices and accessories can be selected based on specific gases. HAXTEC will recommend a complete system to customers after obtaining detailed gas content information.
For hazardous area applications, sensors can be remotely installed in the hazardous area and connected to electrical devices within the safe area through a Zener safety barrier. In addition, when used in more hazardous areas, electrical devices can also be installed in EExd explosion-proof enclosures and non-contact calibration can be performed through the remote control keypad.
`Thermal Conductivity Gas Analyzer
· Reliable technology from professional gas analyzer manufacturer
· Hazardous Area Options
· Corrosion-resistant Gas Sensor
· Loss-free Remote Sensor
Applications
· Hydrogen (H2) Analysis · Gas Mixing
· Synthesis Gas · Nitriding Furnace
· Hydrogen Production · Fuel Cell Research
· Helium (He) Recovery
Optional Measuring Ranges
Measured Component
Minimum Measuring Range
Information
H2 mixed with N2 or Air
0-2 Vol%
Refer to datasheet for other components and ranges
The high measuring range 98-100 Vol% corresponds to the range 0-2 Vol%
H2 mixed with O2
He mixed with N2
Ar mixed with N2
O2 mixed with H2
Customizable for other gases
Custom ranges available: 0-20%, 80-100%, 0-100%
Stability
<1% F.S.D.
Accuracy
±2% F.S.D. (depends on range and gas)
Repeatability
Response Time
T90: 5 seconds (typical)
Sample Flow Rate
100-300 mL/min for optimal performance
Sample Pressure
Nominal atmospheric pressure, adjustable according to exhaust pressure
Sample Connection
Inlet & Outlet: fits 0.25 inch (or 6 mm)
Display
LCD, 2 or 4 lines alphanumeric characters
Analog Output
4-20 mA (user programmable)
Alarm Outputs
2 alarms: each configurable as Off, High or Low by user
Operating Ambient Temperature Range
-5℃ to +40℃
Storage Temperature Range
-10℃ to 55℃
Power Supply
220/240 VAC, 50/60 Hz, power consumption 12 VA