The K1550 series online analyzer is particularly suitable for measuring the VOL% level of a gas in binary or pseudo binary mixtures.
The K1550 series analyzer is particularly suitable for measuring the VOL% level of a gas in binary or pseudo binary mixtures.
We can also provide multi range measuring instruments according to customer requirements, and all models offer a measurement range of 0-100%. Depending on the measured gas and background gas conditions, ranges of 0-20% and 80-100% can also be selected. The hydrogen analyzer can also provide a measurement range of 0-5%.
Applications
· Hydrogen Analysis · Nitriding Furnace
· Syngas · Gas Mixing
· Gasification · Fuel Cell Research
· Helium Recovery
· Sulfur Hexafluoride (SF6) in High Voltage Switchgear
Technical Specifications
Optional Measuring Ranges
(Depends on background measurement gas)0-20%, 80-100%, 0-100% (for most gases)0-5% for most hydrogen concentration measurementsConsult Hanks for other gases
Stability
<1% F.S.D. per month
Accuracy
±2% F.S.D. (Depends on measuring range and gas type)
Response Time
T90: 5 seconds (typical)
Sample Flow Rate
100-300 mL/min for optimal performance
Sample Gas Pressure
Nominal atmospheric pressure, adjustable according to exhaust pressure
Sample Gas Connection
Inlet & Outlet: Compression fittings for hoses with 0.25 inch (or 6 mm) outer diameter
Display
LCD with 2 or 4 lines of alphanumeric characters
Analog Output
4-20 mA (User programmable)
Alarm Outputs
2 alarms: Each alarm can be configured by user as Off, High or Low12 VA
Home
Phone
Message
Top